- Investigate the potential
of NMR spectrometry in process analysis
- Develop a low-field
high-resolution NMR spectrometer for use in process analysis
- Evaluate potential
of at-line and on-line NMR spectrometry using example applications
- Develop signal and
data processing methods to analyse overlapping signals
- Compare NMR spectrometry
with other process analytical techniques
- Conduct a plant-based
evaluation of technique
Achievements
1 Instrumentation
A low-field
NMR spectrometer (operating frequency of 29 MHz for 1H)
has been developed by Resonance Instruments (Witney, UK). The homogeneity
of the magnet was increased through improvements in magnet design
and the addition of shim coils. A lock channel was also employed
to improve the stability of the magnetic field. Using the instrument,
it is possible to acquire 1H, 19F and 31P
NMR spectra with sufficient resolution to enable chemical shift
information to be obtained. There are holes in the top and bottom
of the magnet box, and the probes employ a removable stop at the
base to allow on-line measurements to be performed.
2 Applications
At-line 1H
NMR spectrometry
- determination of strong
acid concentration
- determination of alkyl
chain length in nonyl phenol ethoxylates
- compositional analysis
of samples from a methacrylamide process
- qualitative analysis
of samples from a benzene production plant
- qualitative and quantitative
analysis of polyether polyols
- quantitative analysis
of the esterification reaction of crotonic acid and 2-butanol
- qualitative analysis
of the esterification reaction of itaconic acid and 1-butanol
At-line 19F
NMR spectrometry
- quantitative analysis
of the dehydroxylation of tetrafluorohydroquinone
- quantitative analysis
of the aminolysis of 4-fluorobenzoyl chloride
- quantitative analysis
of an Avecia process
On-line NMR spectrometry
- quantitative analysis
of the esterification reaction of crotonic acid and 2-butanol
3 Data Analysis
- PLS analysis of spectral
data
- Analysis of low-field
NMR signals in the time domain using the continuous wavelet transform
- Analysis of low-field
NMR signals in the time domain using a modification of the generalised
rank annihilation method (FID-GRAM)
- Analysis of low-field
NMR signals in the time domain using a modification of the direct
exponential curve resolution algorithm (FID-DECRA)
4 Deliverables
- Six IMB reports have
been issued and a further six are in preparation
- One paper has been
published, one paper is in press, one paper has been submitted
for publication and a further six papers are in preparation
- Oral presentations
have been given at a number of conferences including IFPAC 2000
(invited) and 2001, FACSS 1999, CAC 2000 and Pittcon 2001. An
invited presentation will also be given at RSC Annual Conference
2001.
5 Objectives versus
Achievements
With the exception of
the final objective, all the objectives were realised during the
project.
- A thorough review
of the literature was prepared (CPACT project 3 IMB report, document
ref. 98/P3/1), which was subsequently submitted to ‘Analyst’ for
publication (A. Nordon, C.A. McGill and D. Littlejohn, Analyst,
2001, 126, 260).
- A low-field high-resolution
NMR spectrometer was developed by Resonance Instruments.
- A number of samples
were obtained from the member companies to assess the potential
use of at-line and on-line NMR spectrometry. However, the model
esterification reactions conducted within CPACT project 1 were
also used to evaluate on-line NMR spectrometry.
- Low-field NMR signals
were analysed in the frequency domain using standard chemometric
techniques. However, methods for analysis of signals in the time
domain were developed in collaboration with CPACT project 2 (wavelet
transform) and MCEC (FID-GRAM and FID-DECRA).
- Comparisons of NMR
spectrometry with other process analytical techniques, such as
NIR, Raman and UV-visible spectrometries, were made using the
model esterification reactions conducted within CPACT project
1. The reactor facility at University of Strathclyde was used
extensively for this purpose.
- Participation in plant-based
trials was sought from the CPACT member companies, but this objective
was not realised.
Deliverables
Review of Flow and
Process NMR document
Document
Ref: 98/P3/1
Issued:
13 October 1998
Executive Summary
NMR spectrometry
is a widely used analytical technique for structural elucidation
and identification of chemical species in the analysis of many different
materials including organic chemicals, inorganic complexes and large
biological molecules. In this context, the instrumentation used
comprises a high-field superconducting magnet (typically 200-600
MHz), and so detailed chemical shift and coupling constant information
may be extracted from the high-resolution spectra obtained. At the
moment NMR spectrometry is not commonly used in process analytical
chemistry. Indeed, NMR spectrometry has been mentioned for the first
time in the most recent of a number of reviews of process analytical
chemistry, but only then within a miscellaneous techniques section.
This is partly due to NMR techniques being rather complicated and
the signals obtained can be difficult for non-NMR specialists to
understand. NMR instruments can be expensive and the technique is
much less sensitive than others, e.g. IR spectroscopy.
However, NMR spectrometry
has the potential of being a very useful technique in process environments
as it is non-destructive and does not require the measurement probe
to be inserted into the process liquors, which avoids fouling. Another
problem that was faced when NMR spectrometry was initially applied
in at-line or on-line process analysis was that high-field laboratory
instruments were simply moved into the plant. Process operators
found problems in the calibration and maintenance of these instruments.
Recently, small, dedicated low-field (<60 MHz) NMR systems, based
on permanent magnet technology, have been developed, which in turn
has led to more reports of at-line and on-line process applications.
As interest in the use
of NMR spectrometry in process applications is increasing, a review
of the subject has been prepared. The report covers various aspects
of the technique that are relevant to NMR analysis of static or
flowing materials (e.g. foods, chemicals and polymers). The review
begins with a discussion of the factors that affect NMR measurements
of flowing samples. This is followed by sections on low-field NMR
applications, on-line process applications and the use of magnetic
resonance imaging in industry. Brief details are given on current
commercial instrumentation. The final section describes chemometric
and data analysis procedures used to acquire chemical and process
information from NMR measurements, especially when overlap of resonances
occur, as is often encountered in low-field NMR spectra.
The Effect of Resolution
and Signal-to-noise Ratio of Simulated Spectra on the Predictive
Capabilities of a PLS Calibration Model
Document
Ref: 99/P3/1
Issued:
29 September 1999
Summary
Techniques
that are used, or are of potential use, in process analysis tend
to yield spectra which are poorly resolved and less easy to interpret
than signals produced by laboratory-based techniques. To extract
the desired information from overlapping signals, chemometric procedures,
such as partial least squares (PLS), are commonly used. In this
study, spectra have been simulated for a two component mixture in
order to examine the factors that affect the predictive ability
of PLS. The factors considered include resolution, signal-to-noise
ratio and lineshape. Although a decrease in predictive accuracy
was observed for poorly resolved signals, the signal-to-noise ratio
was found to have the greatest influence on accuracy. Whether the
lineshape was Gaussian or Lorentzian had no effect on the predictive
ability of PLS. It should be possible to use PLS to predict concentrations
with a high degree of accuracy from data acquired using a wide range
of analytical techniques, irrespective of line shape and resolution,
if the signal-to-noise ratio is of a sufficient level.
Potential Applications
of At-Line Process Analysis by 1H NMR Spectrometry at 28.8 MHz
Document
Ref: 99/P3/2
Issued:
15 October 1999
Summary
The development
version of a bench-top low-field NMR spectrometer, operating at
28.8 MHz for 1H, has been evaluated for use in process
analysis. The instrument incorporates a permanent magnet, and is
smaller and more robust than high-field NMR spectrometers operating
at 200-600 MHz for 1H. Areas of potential application
assessed include determination of acid concentration, individual
component concentration in hydrocarbon mixtures and alkyl chain
length, and the monitoring of a batch reaction.
It has been demonstrated
that acid concentration can be determined with a precision of
+ 0.2 M using
low-field NMR, except where the level of iron (III) (and/or other
paramagnetic species) is in excess of 500 µg mL-I. Through
the use of NMR spectrometry in combination with partial least squares
(PLS) calibration, the composition of multi-component mixtures can
be analysed quantitatively. The results obtained for mixtures of
cyclohexane, hexene and toluene were of a comparable degree of accuracy
to those obtained using NIR spectrometry. It was also possible to
determine the concentration of individual species present in an
esterification reaction, thus enabling the extent of reaction to
be monitored. The ester concentrations predicted from the NMR spectra
were in good agreement with the concentrations obtained using GC.
Qualitative analysis of samples taken at various stages in a benzene
production process was also possible using low-field 1H
NMR spectrometry. The removal of non-aromatic C6 species, at different
stages in the process, was identified. The degree of ethoxylation
in a range of nonylphenol ethoxylate samples was determined from
integration of signals present in the low-field 1H NMR
spectrum and the results were in good agreement with the values
obtained using high-field 1H NMR spectrometry
At present, the precision
of the results obtained by low-field process NMR spectrometry is
limited by the instrumentation, although accuracy is comparable
to both NIR and GC. However, developments are currently underway
in this area, which should then enable both the accuracy and the
precision of analyses performed with the Resonance Instruments'
spectrometer to rival that of more established process analytical
techniques.
Consideration of the
effect of flow for on-line measurements by NMR spectrometry at 28.8
MHz
Document
Ref: 99/P3/3
Issued:
15 October 1999
Summary
High- field
NMR spectrometers found in most laboratories are unsuitable for
the rigours of a process environment. However, on-line process NMR
may be feasible through the use of low-field instrumentation based
on a permanent magnet, incombination with chemometric techniques.
In the work described here, a bench-top low-field NMR spectrometer
operating at 28.8 MHz for 1H has been used to monitor
samples on-line. The decrease in signal intensity observed with
increasing flow rate has been reduced by designing a flow cell that
incorporates a premagnetisation region. Concentrations of toluene,
cyclohexane and hexene in streams flowing at 25 mL min-1
were determined with a similar degree of accuracy to that achieved
for static samples when the premagnetisation flow cell was used
and the 1H spectra were analysed with PLS calibration.
Using a simple sampling
loop, it has also been possible to acquire on-line 1H
NMR spectra of an esterification reaction conducted in a 5 L batch
reactor. Through the use of suitable calibration models and a more
advanced sampling loop, it should be possible to analyse quantitatively
batch reactions with on-line NMR spectrometry. At present, the precision
of the results obtained by process NMR is limited by the instrumentation.
However, developments are currently underway in this area, which
should then improve both the accuracy and precision of NMR at low
frequency for both static and flowing samples.
Preliminary investigation
of quantitative process analysis by low-field 19F NMR
spectrometry using a model reaction
Document
Ref: 99/P3/4
Issued:
20 December 1999
Summary
The paper describes
the use of a small, robust spectrometer with a permanent magnet
for the measurement of 19F NMR spectra of mixtures of
tetrafluorohydroquinone, 1,2,4,5-tetrafluorobenzene and 2,3,5,6-tetrafluorophenol.
Partial least squares regression and variable selection were used
to determine the concentration of the three analytes present at
levels up to 10% m/v. It was not necessary to use the full spectrum
to obtain the optimum standard errors of prediction (SEP). When
10 Hz line broadening was applied to the data, the signal-to-noise
ratio increased while the resolution decreased. In general, this
lowered the SEP values and detection limits were obtained for the
three components, emphasising the importance of acquiring good signal-to-noise
data. The SEP values obtained with 10 Hz line broadening were 0.038%
m/v, 0.074% m/v and 0.164% m/v for tetrafluorohydroquinone, 1,2,4,5-tetrafluorobenzene
and 2,3,5,6-tetrafluorophenol, respectively. The detection limits
obtained for tetrafluorohydroquinone, 1,2,4,5- tetrafluorobenzene
and 2,3,5,6-tetrafluorophenol were 0.01 % m/v, 0.03% m/v and 0.11%
m/v, respectively. The use of lst and 2nd
derivatives of the 19F NMR data did not
improve the quality of the calibration. Although the accuracy of
the NMR results was
good, the precision of the measurements (1-5%) was poorer than that
obtained by near-infrared
spectrometry (0.1-3%). This was due to limitations in the NMR development
instrument which are presently being removed.
Quantitative Analysis of Low-field NMR Signals
in the Time Domain
Document Ref: 01/P3/1
Issued: 26 February 2001
Quantitative analysis
of low-field NMR signals in the time domain
Alison Nordon,1
Paul J. Gemperline,2 Colin A. McGill1 and
David Littlejohn,1
1 Department of Pure
and Applied Chemistry/CPACT, University of Strathclyde, 295 Cathedral
Street, Glasgow, G1 1XL, UK.
2
Department of Chemistry/MCEC, East Carolina University, Greenville,
NC, USA.
Background
Two methods,
based on adaptations of the Generalised Rank Annihilation Method
(GRAM) and the Direct Exponential Curve Resolution Algorithm (DECRA),
have been developed which allow direct quantitative analysis of
the time domain signal, i.e. free induction decay (FID), without
the need for construction of calibration models.
Experimental
The two methods,
termed FID-GRAM and FID-DECRA, were written in Matlab version 5.3
(Mathworks Inc., MA, USA). The performance of each method was investigated
using simulated data. FID-DECRA was also used to analyse 19F
NMR data, which was obtained using a NMR spectrometer (operating
frequency of 29 MHz for 1H) manufactured by Resonance
Instruments. Spectral data, i.e. after Fourier transformation, were
also analysed using partial least squares (PLS) for comparison.
Results
Both FID-GRAM
and FID-DECRA can be used to analyse NMR signals without the need
for the application of a phase correction to the data. However,
as FID-DECRA does not require a reference sample for quantification,
this procedure is preferable. Application
of FID-DECRA to 19F NMR data gave concentrations comparable
to those derived from PLS analysis of spectral data (typical spectrum
shown in figure). The table shows the results obtained for tetrafluorohydroquinone.
However, only a single calibration sample was employed with FID-DECRA,
whereas ten calibration samples were required for PLS.
Conclusions
As FID-DECRA
does not require phase correction nor construction of multivariate
calibration models, it is likely to be particularly useful in process
NMR spectrometry.
Determination of ethylene
oxide content of polyether polyols by low-field 1H NMR
spectrometry
Document Ref:01/P3/3
Issued: 27 September 2001
Background
Methods have been developed and compared for the analysis of
a glycerol-based polyether polyol using a low-field, medium-resolution
NMR spectrometer.
Experimental
Twenty samples of a glycerol-based polyether polyol with an
average relative molecular mass of 5000 and an EO content of 14.8-15.9 %
w/w were analysed at-line using a low-field NMR spectrometer (29 MHz
for 1H; Resonance Instruments). Two methods of obtaining
signal areas were compared, spectral integration (frequency domain)
and a direct exponential curve resolution algorithm for FID analysis
(time domain), known as FID-DECRA. Both methods were used to calculate
directly the EO content of the samples and the results were compared
to those of high-field 13C NMR spectrometry.
Results
FID-DECRA offered advantages over spectral integration as phase
correction was not required and a 2-fold lower RSD was achieved.
Even so, direct analysis gave a negative bias in the EO concentration
(c.f. the 13C NMR results).
|
EO concentration/(% w/w)
|
|
high-field 13C NMR
|
low-field 1H NMR
|
|
spectral analysis
|
FID-DECRA analysis
|
|
15.1
|
15.2 ± 0.23
|
15.1 ± 0.06
|
|
15.4
|
15.5 ± 0.23
|
15.3 ± 0.03
|
When
a univariate calibration model, based on the 13C NMR
results for some of the samples, was used (see table) more accurate
(average % error of 0.5 %) and precise (average RSD of 0.6 %)
concentrations of EO were obtained with FID-DECRA.
Conclusions
FID-DECRA has accuracy and precision advantages over spectral
integration for the at-line determination of % EO content of polyether
polyols. The procedure is also easier to automate, which desirable
in process analysis.
Comparison of calibration
methods for the monitoring of a fluorobenzene batch reaction using
low-field 19F NMR, 1H NMR, NIR and Raman spectrometries
Document Ref: 01/P3/4
Issued: 27 September 2001
Background
The suitability of different process spectrometry techniques has
been assessed, in terms of calibration requirements, accuracy and
precision, for the at-line monitoring of the sulphonation of fluorobenzene.
Experimental
Mixtures of fluorobenzene, 4-fluorobenzenesulphonyl chloride and
4-fluorophenyl sulphone were prepared in dichloromethane to represent
the reaction shown. Samples were analysed by low-field 1H
and 19F NMR (Resonance Instruments), NIR (Bomem) and
Raman (Kaiser) spectrometries.
Results
|
4-fluorobenzene
sulphonyl chloride/
mol dm-3
|
NIR (PLS)/
mol dm-3
|
19F
NMR (univariate)/
mol dm-3
|
1H
NMR (PLS)/
mol dm-3
|
|
0.22
|
0.24 ±
0.002
|
0.22 ±
0.018
|
-0.28 ±
0.117
|
|
2.16
|
2.22 ±
0.005
|
2.26 ±
0.006
|
1.87 ±
0.305
|
Partial least squares
(PLS) calibration was required to analyse the spectra obtained by
NIR spectrometry and low-field (29.1 MHz) 1H NMR
spectrometry. The low-field (27.4 MHz) 19F NMR spectra
contained well-resolved signals for the three fluorine-containing
compounds and univariate calibration was adequate. The Raman spectra
of two of the compounds exhibited fluorescence and so this technique
was not considered suitable for monitoring the reaction. The most
accurate and precise results were obtained using PLS analysis of
NIR data and univariate analysis of 19F NMR data. Results
of poor accuracy were obtained using PLS analysis of 1H
NMR data as spectra for two of the compounds were too similar.
Conclusions
NIR and 19F NMR spectrometries are suitable for the monitoring
of the sulphonation of fluorobenzene. Analysis of the NIR data required
construction of PLS calibration models, whereas univariate calibration
was satisfactory for 19F NMR spectra. Hence, low-field
19F NMR spectrometry may offer a simpler alternative
method of analysis to NIR spectrometry for the monitoring of fluorocarbon
processes.
|